Rg. Rudnitsky et al., Rapid biochemical detection and differentiation with magnetic force microscope cantilever arrays, SENS ACTU-A, 83(1-3), 2000, pp. 256-262
We have tested a novel method of rapid bead-on-surface detection using a ma
gnetic force microscope (MFM) with a ferromagnetic tip. This technique's re
liance on mechanical sensors should allow for high speed, throughput and se
nsitivity. In this report, we demonstrate that the fundamental physics allo
ws for the rapid detection of a single bead bound to a surface by magnetic
force sensing. (C) 2000 Elsevier Science S.A. All rights reserved.