Evaluation of different SIA sample-buffer configurations using a fluoride-selective membrane electrode as detector

Citation
Jkf. Van Staden et al., Evaluation of different SIA sample-buffer configurations using a fluoride-selective membrane electrode as detector, TALANTA, 52(1), 2000, pp. 3-11
Citations number
19
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
TALANTA
ISSN journal
00399140 → ACNP
Volume
52
Issue
1
Year of publication
2000
Pages
3 - 11
Database
ISI
SICI code
0039-9140(20000531)52:1<3:EODSSC>2.0.ZU;2-0
Abstract
The peak profiles of four different buffer-sample SIA configurations, e.g. buffer-sample; sample-buffer; buffer-sample-buffer and sample-buffer-sample with the last two in the sandwich mode were evaluated with a fluoride-sele ctive membrane electrode as detector. The best response characteristics and peak shapes as well as recovery and precision values were obtained for the buffer-sample configuration. For low concentration levels, sandwich SIA co nfigurations are more suitable, when optimum buffer and pH are used. The ut ilisation of a cheap electrolyte with a minimum consumption of buffer and a cheap, robust instrumentation made the SIA system suitable for on-line det ermination of fluoride. (C) 2000 Elsevier Science B.V. All rights reserved.