An enzymatically modified ammonium ion-selective electrode has been applied
for the determination of urea in spent dialysate. The biosensor has been u
sed in a simple flow-injection analysis (FIA) system. The system enables on
e to perform over 25 dialysate urea nitrogen (DUN) determinations per hour.
The interferences from other components of posthemodialysis fluid were eli
minated by simultaneous measurements with non-modified enzymatically ion-se
lective electrode. It is possible to use both the sensors in a simplified d
ifferential potentiometric system. The results of DUN determination using t
he biosensor/FIA system and a conventional method of urea determination wer
e comparable. The presented analytical system can potentially find wider bi
omedical application in the monitoring of hemodialysis progress. (C) 2000 E
lsevier Science B.V. All rights reserved.