H. Du Plessis et Jf. Van Staden, Determination of sulphuric acid in process effluent streams using sequential injection titration, TALANTA, 52(1), 2000, pp. 83-90
Sulphuric acid in process effluent streams from an electrorefining copper p
lant was analysed with a sequential injection (SI) titration system using s
odium hydroxide as titrant. In the proposed SI titration system a base titr
ant, acid analyte and base titrant zone were injected sequentially into a d
istilled water carrier stream in a holding coil and swept by flow reversal
through a reaction coil to the detector. The base zones contained bromothym
ol blue as indicator and the endpoint was monitored spectrophotometrically
at 620 nm. The influence of carrier stream flow rate, acid and base zone vo
lumes and titrant concentration on the linear range of the method was studi
ed to obtain an optimum. A linear relationship between peak width and logar
ithm of the acid concentration was obtained in the range 0.006-0.178 mol l(
-1) of H2SO4 for a NaOH concentration of 0.002 mol l(-1). The results obtai
ned for the SI titration of process samples were in good agreement with a s
tandard potentiometric method with an RSD < 0.75% and a sample frequency of
23 samples h(-1). (C) 2000 Elsevier Science B.V. All rights reserved.