I. Feldmann et al., PERFORMANCE-CHARACTERISTICS OF INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY WITH HIGH-MASS RESOLUTION, Journal of analytical atomic spectrometry, 9(9), 1994, pp. 1007-1014
A laboratory prototype is described of a new ICP-MS system with high m
ass resolution, based on a double focusing magnetic field analyser, wh
ich has been especially designed for application in elemental analysis
. The performance characteristics of this instrument are summarized co
mparing operation in a low resolution mode (R = 300) and a high resolu
tion mode (R = 3000) with reference to common quadrupole instruments w
ith (nearly) unit resolution. Most of the typical spectral interferenc
es can be resolved. Detection limits were achieved of one order of mag
nitude and more below those obtained with quadrupole instruments. Seve
ral operational features of the instrument are described.