Abstract
Accurate estimation of resultant joint forces and moments from gait an alysis data depends heavily on the accuracy of measurement of the grou nd reaction force (GRF). Typically, multicomponent force platforms are used to measure the components of the GRF and the position of the cen tre of pressure. A new rig, designed to allow safe and quick static te sting of the vertical component and centre of pressure outputs of such force platforms, is described. This rig has been used to test the acc uracy of AMTI force platforms and was found to reduce dramatically the testing time, The results of the tests are also given.