Dry calibration procedure of electromagnetic flowmeter for open channels

Authors
Citation
A. Michalski, Dry calibration procedure of electromagnetic flowmeter for open channels, IEEE INSTR, 49(2), 2000, pp. 434-438
Citations number
11
Categorie Soggetti
Instrumentation & Measurement
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
ISSN journal
00189456 → ACNP
Volume
49
Issue
2
Year of publication
2000
Pages
434 - 438
Database
ISI
SICI code
0018-9456(200004)49:2<434:DCPOEF>2.0.ZU;2-P
Abstract
This paper presents a simple approach to a dry calibration procedure for th e electromagnetic flowmeter for both conducting and nonconducting open-flow channels. The author focuses on two main problems: variable mater level in the measurement zone and variable water or ground conductivity and their i nfluence on a flow signal. Based on the existing numerical model of electro magnetic flowmeter, the author describes the main idea of transfer function calculation with a few exemplary results. As an approximation instrument f or a final form of correction function, the sigmoidal two-layer neural netw ork has been applied.