A continuous motional series resonant frequency monitoring circuit and a new method of determining Butterworth-Van Dyke parameters of a quartz crystal microbalance in fluid media

Citation
A. Arnau et al., A continuous motional series resonant frequency monitoring circuit and a new method of determining Butterworth-Van Dyke parameters of a quartz crystal microbalance in fluid media, REV SCI INS, 71(6), 2000, pp. 2563-2571
Citations number
69
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
71
Issue
6
Year of publication
2000
Pages
2563 - 2571
Database
ISI
SICI code
0034-6748(200006)71:6<2563:ACMSRF>2.0.ZU;2-W
Abstract
A new electronic design for continuous monitoring of the motional series re sonant frequency of a quartz crystal microbalance either in vacuum or in si ngle-face contact with a fluid medium is presented. The motional resistance is directly measured and a simple method for measuring L and C parameters of the Butterworth-Van Dyke equivalent circuit is described. The results ar e in agreement with impedance analyzer measurements. (C) 2000 American Inst itute of Physics. [S0034-6748(00)04806-1].