P. Liu et al., On-line preconcentration and separation of platinum using thiourea modified silica gel with microwave assisted desorption for FAAS determination, ANALYST, 125(6), 2000, pp. 1205-1209
A flow-injection on-line preconcentration and separation system with microw
ave assisted desorption is proposed for the efficient flame atomic absorpti
on spectrometric determination of trace amounts of Pt. A microcolumn packed
with thiourea modified silica gel placed in a microwave oven was used to c
ollect Pt from sample solutions. 2% thiourea was used to elute the adsorbed
Pt into the nebulizer of the spectrometer with microwave irradiation. Opti
mization studies were carried out on the volume of cooling water, heating t
ime of the microwaves, sample acidity, concentration of eluent etc. Common
co-existing ions did not interfere with the determination of Pt(IV). The RS
D of determination of 1.0 mu g ml(-1) was < 4% with 120 s of preconcentrati
on at a sampling flow rate of 2.5 ml min(-1). The limit of detection, defin
ed as three times the standard deviation of blank (3 sigma), was 0.060 mu g
ml(-1). The results obtained by the proposed method for three types of met
allurgical samples, a secondary nickel alloy, anode slime and a CoCl2 solut
ion, were in good agreement with the recommended values.