Cm. Sutton et al., Conventional pressure balances as reference standards for pressures in therange 10 Pa to 10 kPa, METROLOGIA, 36(6), 1999, pp. 517-520
Two methods are presented for using conventional pressure balances in the p
ressure range 10 Pa to 10 kPa. Both these methods overcome the inability of
a pressure balance to generate directly pressures below a limit determined
by the mass of the floating element. The first method uses twin pressure b
alances, with the desired low pressure established above one pressure balan
ce in terms of a small added mass on the other pressure balance. This metho
d is shown to be capable of realizing low absolute pressures with a standar
d uncertainty below 2 mPa + (9 x 10(-6))p, when operating the pressure bala
nces at 100 kPa. The second method uses one pressure balance in conjunction
with a high-resolution pressure gauge. A calculable low pressure is establ
ished above the pressure balance either by reference to an added mass on th
e pressure balance or by a change in the pressure gauge reading. A 100 kPa
pressure balance is used in combination with a commercial high-precision ba
rometric pressure gauge, modified to improve the temperature control of its
sensor. The standard uncertainty associated with this implementation is sh
own to be about 14 mPa+(9 x 10(-6))p.