Quality assurance of automated gamma-ray spectrometric analysis

Citation
D. Glavic-cindro et al., Quality assurance of automated gamma-ray spectrometric analysis, APPL RAD IS, 53(1-2), 2000, pp. 237-241
Citations number
6
Categorie Soggetti
Multidisciplinary
Journal title
APPLIED RADIATION AND ISOTOPES
ISSN journal
09698043 → ACNP
Volume
53
Issue
1-2
Year of publication
2000
Pages
237 - 241
Database
ISI
SICI code
0969-8043(200007/08)53:1-2<237:QAOAGS>2.0.ZU;2-O
Abstract
Fully automatic gamma-ray spectrometric analysis procedures perform complet e processing of the spectrum without intervention of the operator. In order to maintain the reliability of the final results the analysis checks the i ntermediate results automatically. When a disagreement is identified by suc h a check the uncertainty of the intermediate results is increased in order to accommodate the disagreement. The increased uncertainty is propagated i nto the uncertainty of the final results in order to take into account the disagreement. This approach was implemented in Canberra's Genie ESP gamma-r ay spectrometry package for examining the results of the peak analysis. In addition to this intermediate check also a-posteriori checks of the final r esults can be performed by statistical analysis. Such analysis shows whethe r the results are under statistical control and can discover sources of var iability which are not taken into account in the uncertainty budget. (C) 20 00 Elsevier Science Ltd. All rights reserved.