Y. Yamanouchi et al., OPTIMIZED FIRST-PHASE TILT IN PARALLEL-SERIES BIPHASIC WAVE-FORM, Journal of cardiovascular electrophysiology, 8(6), 1997, pp. 649-657
Introduction: A biphasic defibrillation waveform can achieve a large s
econd phase leading-edge voltage by a ''parallel-series'' switching sy
stem. Recently, such a system using two 30-mu F capacitances demonstra
ted better defibrillation threshold than standard waveforms available
in current implantable devices, However, the optimized tilt of such a
''parallel-series'' system had not been defined. Methods and Results:
Defibrillation thresholds were evaluated for five different biphasic '
'parallel-series'' waveforms (60/15 mu F) and a biphasic ''parallel-pa
rallel'' waveform (60/60 mu F) in 12 anesthetized pigs. The five ''par
allel-series'' waveforms had first phase tilts of 40%, 50%, 60%, 70%,
and 80% with second phase pulse width of 3 msec, The ''parallel-parall
el'' waveform had first phase tilt of 50% with second phase pulse widt
h of 3 msec. The defibrillation lead system comprised a left pectoral
''hot can'' electrode (cathode) and a right ventricular lead (anode),
The stored energy at defibrillation threshold of the ''parallel-series
'' waveform with first phase tilts of 40%, 50%, 60%, 70%, and 80% was
7.0 +/- 2.1, 6.1 +/- 2.8, 6.8 +/- 2.8, 7.2 +/- 2.9, and 8.4 +/- 3.1 J,
respectively, The stored energy of the ''parallel-series'' waveform w
ith a 50% first phase tilt was 16% less than the nonswitching ''parall
el-parallel'' waveform (7.3 +/- 2.8 J, P = 0.006). Conclusions: A firs
t phase tilt of 50% maximized defibrillation efficacy of biphasic wave
forms implemented with a ''parallel-series'' switching system. This op
timized ''parallel-series'' waveform was more efficient than the compa
rable ''parallel-parallel'' biphasic waveform having the same first ph
ase capacitance and tilt.