Contrast-enhanced MR angiography: The effects of k-space truncation on luminal representation in a carotid artery phantom model

Citation
Er. Melhem et al., Contrast-enhanced MR angiography: The effects of k-space truncation on luminal representation in a carotid artery phantom model, AM J NEUROR, 21(6), 2000, pp. 1028-1031
Citations number
9
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Neurosciences & Behavoir
Journal title
AMERICAN JOURNAL OF NEURORADIOLOGY
ISSN journal
01956108 → ACNP
Volume
21
Issue
6
Year of publication
2000
Pages
1028 - 1031
Database
ISI
SICI code
0195-6108(200006/07)21:6<1028:CMATEO>2.0.ZU;2-F
Abstract
Using carotid bifurcation phantom models with different degrees of stenoses , we evaluated the accuracy of vessel lumen representation on MR images obt ained from the inverse Fourier transform of different k-space percentages. Our results show that the lower thresholds of truncated k-space sampling ar e dictated by the severity of luminal narrowing. The defined thresholds may help improve efficiency of 3D MR imaging of the carotid arteries while mai ntaining adequate luminal representation.