Aa. Ensafi et Ma. Naeeni, Determination of copper by an adsorption differential pulse stripping method with naphthol derivative, ANAL LETTER, 33(8), 2000, pp. 1591-1601
A simple, sensitive and selective method is proposed for the determination
of copper(II) by adsorption-differential pulse shipping method. This method
is based on the selective accumulation of the complexes of Cu(II) with 1-(
2-pyridylazo)-2-Naphthol and then reduction of the complex on a HMDE. The r
eduction current of the complex is about 0.0V vs. Ag/AgCl reference electro
de at pH=4.0. The influences of various experimental parameters on the curr
ent peak were completely studied. The calibration graph was linear up to 50
.0 ng/ml for a deposition time of 60 sec. The relative standard deviation w
as 2.3% (n=5) for Cu(II) concentration of 10.0 ng/ml, with a limit of detec
tion of 0.2 ng/ml. The influence of potentially interfering ions was comple
tely studied. The method has been applied for the determination of Cu(II) i
n wafer samples.