Sa. Wade et al., Strain-independent temperature measurement by use of a fluorescence intensity ratio technique in optical fiber, APPL OPTICS, 39(18), 2000, pp. 3050-3052
The strain sensitivity of the fluorescence intensity ratio temperature-sens
ing technique has been measured to be (2 +/- 3) x 10(-4)%/mu epsilon in Yb3
+-doped fiber, implying a temperature-to-strain cross sensitivity of(2 +/-
3) x 10(-4) degrees C/mu epsilon. The near-zero strain sensitivity means th
at this optical-fiber sensor technique is well. suited for temperature meas
urement in strain-affected environments. (C) 2000 Optical Society of Americ
a. OCIS codes: 060.2370, 120.6780, 160.5690, 300.2530.