INVESTIGATION OF ELECTRICAL CONTACTS FOR CD1-XZNXTE NUCLEAR RADIATIONDETECTORS

Citation
A. Burger et al., INVESTIGATION OF ELECTRICAL CONTACTS FOR CD1-XZNXTE NUCLEAR RADIATIONDETECTORS, IEEE transactions on nuclear science, 44(3), 1997, pp. 934-938
Citations number
5
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
44
Issue
3
Year of publication
1997
Part
1
Pages
934 - 938
Database
ISI
SICI code
0018-9499(1997)44:3<934:IOECFC>2.0.ZU;2-G
Abstract
A comparative study of four different metal deposition methods on dete ctor grade Cd1-xZnxTe crystal has been done using current-voltage (I-V ) characterization and Ba-133 gamma spectroscopy. Electroless Au conta ct and thermal evaporated Au contact have been found to yield a smalle r leakage current and a more linear/symmetric I-V curve than that of s puttered Pt and Au contact. An optimum detector performance can be ach ieved by combining the thermal evaporated Au deposition method with a chemical surface treatment consisting of 5%Br-MeOH followed by 2% Br-2 0% lactic acid in ethylene glycol. Low temperature photoluminescence s pectra of Cd1-xZnxTe samples corresponding to these different metalliz ation schemes are also presented and discussed.