V. David et al., Enhanced chemiluminescent determination of chloramphenicol and related nitro compounds by 'on-line' photochemical reaction, ANALYST, 125(7), 2000, pp. 1313-1319
The viability of tandem photochemical reaction-chemiluminescence detection
was studied for a heterogeneous family of nitro compounds using chloramphen
icol as a test substance. The 'on-line' chemical photodegradation of chlora
mphenicol was performed in a flow injection assembly by using a photoreacto
r consisting of a 725 cm x 0.5 mm id piece of PTFE tubing coiled around an
8 W low-pressure mercury lamp. Photodegraded chloramphenicol was detected b
y oxidizing photo-fragments from the parent compound and their subsequent r
eaction with a luminol-Co(II) system. The calibration graph was linear up t
o 3 x 10(-5) mol l(-1) chloramphenicol, the limit of detection was 3 x 10(-
9) mol l(-1), the relative standard deviation was 1.8% for 2 x 10(-6) mol l
(-1) of the drug and the sample throughput was 60 h(-1). The proposed metho
d was used to determine chloramphenicol in pharmaceutical formulations and
its application to other related nitro compounds was studied.