A new cyclodextrin-modified micellar electrokinetic chromatography (CD-MEKC
) method for the enantiomeric separation and determination of the selenoami
noacids selenomethionine and selenoethionine, derivatised with 2,3-naphthal
enedicarboxaldehyde (NDA) to produce cyanobenzoisoindole (CBI) derivatives,
has been developed. Optical resolution was achieved in a fused-silica capi
llary column using 30 mmol l(-1) phosphate/10 mmol l(-1) boric acid (pH 7)
electrophoretic buffer containing a mixed micellar system of achiral sodium
dodecyl sulphate and chiral taurodeoxycholic acid surfactants among with p
-cyclodextrins. Detection was followed by direct UV absorptiometric measure
ments at 230 nm. The concentration influence of the two chiral selectors us
ed (beta-CD and TDC micelles), the effect of applied voltage and the pH of
the electrophoretic buffer on the migration time and resolution have been i
nvestigated. The analytical performance of the method is discussed in terms
of detection limits, linearity of response and precision. The method devel
oped has been applied to the determination of enantiomeric purity in a comm
ercial sample sold as 'pure' L-selenomethionine. (C) 2000 Elsevier Science
B.V. All rights reserved.