Determination of daunorubicin in human urine by capillary zone electrophoresis with amperometric detection

Citation
Q. Hu et al., Determination of daunorubicin in human urine by capillary zone electrophoresis with amperometric detection, ANALYT CHIM, 416(1), 2000, pp. 15-19
Citations number
10
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICA CHIMICA ACTA
ISSN journal
00032670 → ACNP
Volume
416
Issue
1
Year of publication
2000
Pages
15 - 19
Database
ISI
SICI code
0003-2670(20000703)416:1<15:DODIHU>2.0.ZU;2-J
Abstract
A simple, reliable and reproducible method, based on capillary zone electro phoresis with amperometric detection, for the determination of daunorubicin in human urine is described in this paper. A carbon disk electrode used as working electrode. The optimal conditions of separation and detection were pH 7.8 phosphate buffer (0.15 mol/l), 8 kV for the separation voltage and 0.95 V (versus Ag/AgCl, 3 mol/l KCl) for the detection potential. The linea r range was from 2x10(-6) to 2x10(-4) mol/l with the regression coefficient of 0.9995, and the detection limit was 8 x 10(-7) mol/l. The method built in this paper was directly applied to the determination of daunorubicin in the imitation human urine without any other sample pre-treatment except fil tration, and the assay result was satisfactory. (C) 2000 Elsevier Science B .V. All rights reserved.