Studies on the use of commercial capillary gas chromatographic columns as diffusion denuders

Citation
M. Dudek et al., Studies on the use of commercial capillary gas chromatographic columns as diffusion denuders, HRC-J HIGH, 23(6), 2000, pp. 449-454
Citations number
31
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
HRC-JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY
ISSN journal
09356304 → ACNP
Volume
23
Issue
6
Year of publication
2000
Pages
449 - 454
Database
ISI
SICI code
0935-6304(200006)23:6<449:SOTUOC>2.0.ZU;2-Q
Abstract
The possibility of using a piece of gas chromatographic capillary column as a tubular denuder for Isolation and enrichment of organic pollutants prese nt in air was studied. The partition coefficients (K-fs) of typical organic pollutants (benzene, chlorobenzene, toluene) between the gaseous phase and the denuder sorption layer were determined and effects of analyte concentr ation in the gaseous phase and gas flow rate on partitioning were studied. The partition coefficients were found to he constant in the studied range o f concentration and gaseous mixture flow rate. A piece of a capillary colum n coated with polydimethylosiloxane can he used as a tubular denuder.