Jf. Vanstaden et Re. Taljaard, DETERMINATION OF AMMONIA IN WATER AND INDUSTRIAL EFFLUENT STREAMS WITH THE INDOPHENOL BLUE METHOD USING SEQUENTIAL INJECTION-ANALYSIS, Analytica chimica acta, 344(3), 1997, pp. 281-289
A suitable sequential injection analysis (SIA) analyser to monitor wat
er quality in terms of the ammonia content in water and industrial eff
luent streams was developed. The system is fully computerised and is a
ble to monitor ammonia in samples at a frequency of ca. 16 samples per
hour with a relative standard deviation better than 1.8%. The calibra
tion curve is linear between 0 and 50 mg l(-1) which falls within the
required analyte range. The levels of possible interferents in the wat
er and industrial effluent streams were negligible to cause any interf
erence on the proposed system.