Use of an implantable cardioverter defibrillator in a patient with two implanted neurostimulators for severe Parkinson's disease

Citation
R. Tavernier et al., Use of an implantable cardioverter defibrillator in a patient with two implanted neurostimulators for severe Parkinson's disease, PACE, 23(6), 2000, pp. 1057-1059
Citations number
3
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Journal title
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
ISSN journal
01478389 → ACNP
Volume
23
Issue
6
Year of publication
2000
Pages
1057 - 1059
Database
ISI
SICI code
0147-8389(200006)23:6<1057:UOAICD>2.0.ZU;2-4
Abstract
We report a patient with Parkinson's disease treated with two pectorally im planted neurostimulators (NSs) who presented with a life-threatening ventri cular tachyarrhythmia in whom an abdominal ICD was implanted. Testing durin g implantation showed that the NS did not affect the bipolar sensing of the ICD, even when the NSs were set at a frequency of 130 pulses/s with an out put of 5 V and pulse width of 0.21 ms in a bipolar and a unipolar configura tion. The ICD shock, however, did affect both NSs: there was a reset to the output Off state and there was a reset of the electrode polarities.