We report a patient with Parkinson's disease treated with two pectorally im
planted neurostimulators (NSs) who presented with a life-threatening ventri
cular tachyarrhythmia in whom an abdominal ICD was implanted. Testing durin
g implantation showed that the NS did not affect the bipolar sensing of the
ICD, even when the NSs were set at a frequency of 130 pulses/s with an out
put of 5 V and pulse width of 0.21 ms in a bipolar and a unipolar configura
tion. The ICD shock, however, did affect both NSs: there was a reset to the
output Off state and there was a reset of the electrode polarities.