Y. De Deene et al., Artefacts in multi-echo T-2 imaging for high-precision gel dosimetry: I. Analysis and compensation of eddy currents, PHYS MED BI, 45(7), 2000, pp. 1807-1823
In BANG gel dosimetry, the spin-spin relaxation rate, R-2 = 1/T-2, is relat
ed to the radiation dose that has been delivered to the Eel phantom. R-2 is
calculated by fitting the pixel intensities of a set of differently T-2-we
ighted base images. In gel dosimetry for radiotherapy, an accuracy of 5% in
dose and 3 mm spatially, whichever is lower. is the objective. Therefore,
possible sources of artefacts must be considered and dealt with.
To obtain a set of base images a multiple spin-echo sequence is used. Howev
er, in a conventional MR scanner eddy currents will be provoked by switchin
g the imaging gradients. As the eddy currents change in the course of the s
equence, the net magnetization will be affected accordingly. Hence, eddy cu
rrents may have a significant influence on the quantitative R-2 images them
selves as well as on their slice position.
In this study, we report an analysis of the eddy currents as they appear in
the multiple spin-echo sequence. Eddy currents are measured using a freque
ncy shift method resulting in eddy current field maps. The related geometri
cal displacements are obtained by use of a pyramidal phantom. The R-2 versu
s dose relation is determined in the three main directions of the magnet, r
evealing a dependence of the measured R-2 on slice orientation. The time co
urse of eddy currents is then used in a computer simulation to estimate the
effects they produce in the recorded R-2 images. A compensation method for
eddy current effects in multi-echo T-2 mapping is described.