An automated approach to seed assignment for eye plaque brachytherapy

Citation
T. San Koh et al., An automated approach to seed assignment for eye plaque brachytherapy, PHYS MED BI, 45(7), 2000, pp. 1897-1912
Citations number
9
Categorie Soggetti
Multidisciplinary
Journal title
PHYSICS IN MEDICINE AND BIOLOGY
ISSN journal
00319155 → ACNP
Volume
45
Issue
7
Year of publication
2000
Pages
1897 - 1912
Database
ISI
SICI code
0031-9155(200007)45:7<1897:AAATSA>2.0.ZU;2-O
Abstract
Episcleral plaques are commonly used for the treatment of ocular rumours su ch as choroidal melanoma. Treatment planning involves the assignment of see ds to slots on the plaque to achieve a desired dose rate distribution. Seed assignment is rather straightforward if seeds are ordered on demand. Howev er, the assignment task becomes tedious and laborious if the seeds have to be chosen from an existing stock of seeds with different activities. To dar e, this task has usually been performed by a human planner through trial an d error. An algorithm has been developed to automate the task of seed assig nment using a mixed-integer programming method. We also explore ways to sim plify the problem such that the method becomes practical in most facilities . We have tested the method on three randomly chosen clinical cases from ou r past records, to show that the algorithm could yield solutions within a s horter time frame and with less deviation from the desired dose rate distri butions, as compared with the solutions from a human counterpart.