Rapid and accurate determination of NADH by an amperometric sensor with a bilayer membrane consisting of a polyion complex layer and an NADH oxidase layer

Citation
F. Mizutani et al., Rapid and accurate determination of NADH by an amperometric sensor with a bilayer membrane consisting of a polyion complex layer and an NADH oxidase layer, SENS ACTU-B, 65(1-3), 2000, pp. 46-48
Citations number
8
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
SENSORS AND ACTUATORS B-CHEMICAL
ISSN journal
09254005 → ACNP
Volume
65
Issue
1-3
Year of publication
2000
Pages
46 - 48
Database
ISI
SICI code
0925-4005(20000630)65:1-3<46:RAADON>2.0.ZU;2-D
Abstract
An amperometric NADH-sensing electrode was prepared by immobilizing NADH ox idase on a poly-L-lysine/poly(4-styrenesulfonate)-complex-coated electrode. The polyion complex layer was effective for eliminating electrochemical in terferents such as L-ascorbic acid, uric acid and acetaminophen, whereas th e hydrogen peroxide produced through the NADH oxidase-catalyzed reaction pe rmeated rapidly through the layer. This resulted in a rapid response with a low interference level. An alcohol dehydrogenase/NADH oxidase-bienzyme ele ctrode was also prepared for the measurement of alcohol, and for the highly sensitive measurement of NADH, which was based on the substrate recycling with the dehydrogenase-catalyzed reaction. (C) 2000 Elsevier Science S.A. A ll rights reserved.