J. Girouard et al., Multicenter evaluation of the Glucometer Elite XL Meter, an instrument specifically designed for use with neonates, DIABET CARE, 23(8), 2000, pp. 1149-1153
OBJECTIVE - To evaluate the clinical performance of the Glucometer Elite XL
Diabetes Care System in neonatal settings using a multicenter study.
RESEARCH DESIGN AND METHODS - A total of 388 blood specimens from 333 neona
tes were included in the study. A capillary or arterial sample was analyzed
for determination of glucose with the Glucometer Elite XL system by an att
ending trained nurse. Through the same sampling site, a specimen was collec
ted and sent to the laboratory for measurement of plasma glucose, bilirubin
, and hematocrit.
RESULTS - The regression analysis between the results of the Glucometer Eli
te XL system and comparative methods resulted in the following. Glucometer
Elite XL meter = 1.01 X laboratory method + 0.02 mmol/l (n = 388). For the
1.1-4.0 mmol/l plasma glucose range, the regression was Glucometer Elite XL
meter = 1.07 X laboratory method + 0.12 mmol/l (n = 150). A difference plo
t indicated a mean bias of 0.04 mmol/l (95% CI -0.01 to 0.10). No relations
hip was found between meter glucose biases and hematocrit levels (r = 0.10,
P = 0.14). Although a statistically significant correlation existed betwee
n bilirubin levels and the glucose meter biases (r = 0.14, P = 0.005), the
predicted mean biases were of little clinical significance.
CONCLUSIONS - The Glucometer Elite XL system showed a good performance when
used in neonatal settings.