Evaluation of the BIOMIC video reader system for routine use in the clinical microbiology laboratory

Citation
Hk. Geiss et Ue. Klar, Evaluation of the BIOMIC video reader system for routine use in the clinical microbiology laboratory, DIAG MICR I, 37(3), 2000, pp. 151-155
Citations number
18
Categorie Soggetti
Clinical Immunolgy & Infectious Disease",Microbiology
Journal title
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE
ISSN journal
07328893 → ACNP
Volume
37
Issue
3
Year of publication
2000
Pages
151 - 155
Database
ISI
SICI code
0732-8893(200007)37:3<151:EOTBVR>2.0.ZU;2-O
Abstract
The BIOMIC Video Reader System (Giles Scientific, New York, NY, USA) is a s emi-automated AST method that combines disk diffusion testing with automate d reading and data interpretation. We evaluated this system with 497 strain s comprising a total of 5821 drug organism combinations (DOC) from our rout ine diagnostic laboratory. Additionally, we compared the time required of t he manual and the automated method. The overall agreement of interpretative categories of all DOC was 96.1%. However, comparing complete tests the agr eement was only 70.8%. The average time required of the BIOMIC system to co mplete a test was more than twice as long as that of the manual method. Our data suggest that the tested version of the BIOMIC system cannot be recomm ended for routine use in diagnostic laboratories. (C) 2000 Elsevier Science Inc. All rights reserved.