The response of a wireline-type beam monitor was measured for electron beam
s with energies of 28 MeV and pulsewidths of 20 ps from an electron linear
accelerator. The output waveform from the wireline pickup was calculated ac
cording to the general formulation for responses of beam monitors based on
electromagnetic interaction. In the calculations, a transfer function H(ome
ga) and an effective length l(eff) are introduced to express all the effect
s not included in the analysis. The function H(omega) is assumed to be Gaus
sian, and the FWHM of H(omega) and l(eff) is experimentally determined. The
effects of the termination, wireline length, height, and angle between the
wireline and the beam direction are discussed. The test supported the anal
ysis and, for the standard pickup, a charge sensitivity of 20 V/nC and an F
WHM of H(omega) of 55 ps are obtained.