A procedure has been developed for decreasing the sampling interval for pos
itron emission tomographs (PETs). For this procedure to be incorporated int
o a PET, the gamma ray detectors of the PET must provide depth-of-interacti
on (DOI) information for the detection of the gamma ray. Four radial, as we
ll as an angular, addresses were provided for each scintillation detector i
n the tomograph ring for the PET tested. By tilting each detector toward it
s neighbor, the centers of the four radial segments were displaced one-quar
ter of the angular distance between the centers of the untilted detectors.
The utility of the improved sampling has been demonstrated by measuring the
point-spread-function (PSF) for 1 mn sources separated by 4 mm near the ax
is of the PET under test. With the radial data combined for each detector,
the PSF peaks were difficult to identify since each peak contained only 3 s
ampling points. With the radial data combined for the adjacent radial segme
nts, the 12 sampling points clearly delineated each of the two PSF peaks. F
inally, for comparison, the statistical characteristics for PSFs obtained w
ith wobbling detectors were calculated.