Piribedil (PD) ion-selective electrodes have been constructed from poly(vin
yl chloride) matrix membrane containing piribedil-tetraphenylborate (PD-TPB
) as: the electroactive component with dibutylphthalate or dioctylphthalate
as the plasticizing solvent mediator. The electrodes displayed a linear re
sponse over the concentration range 2.0 x 10(-5) to 10(-2) M PD. The workin
g pH ranges of the electrodes were 3.5-6.4 and 3.0-6.0, and the isothermal
coefficients of the cells were 0.00129 and 0.00096 V/degrees C, respectivel
y. The electrodes were used for the determination of the diprotonated PD sp
ecies, the most successful being that based on dioctylphthalate solvent med
iator. The electrodes show a linear response over the concentration range o
f 8.0 x 10(-6) to 10(-2) M PD, with Nernstain slope 30 mV/PD concentration
decade when preconditioned by soaking in distilled water for 30 min. The el
ectrodes exhibit good selectivity for the PD with respect to a large number
of inorganic cations and organic substances of biological fluids. Piribedi
l is determined successfully in pure solutions and in tablets or in biologi
cal fluids using the standard additions and potentiometric titration method
s. The membrane withstood soaking in distilled water for more than 5 months
. (C) 2000 Elsevier Science B.V. All rights reserved.