Hc. Chang et al., Determination of coagulation time in whole blood containing anticoagulant by piezoelectric quartz crystal sensor, SENS ACTU-B, 66(1-3), 2000, pp. 296-298
A piezoelectric quartz crystal (PQC) system possessing an adequate sensitiv
ity to a slight change of viscosity was previously employed to monitor bloo
d coagulation. Here, the activated partial thromboplastin time (aPTT), clot
ting time (CT) for platelet poor plasma (PPP), and whole blood were evaluat
ed using PQC system, the results showed a good agreement with the aPTT resu
lts measured by optical coagulometer (OC). As a result, a good linear relat
ionship for heparin concentration from 0 to 1 unit/ml introduced in both of
the specimens could be obtained in PQC and OC systems. (C) 2000 Elsevier S
cience S.A. All rights reserved.