Intraoperative test stimulation with a modified implantable pulse generator in deep brain stimulation

Citation
Jk. Krauss et T. Pohle, Intraoperative test stimulation with a modified implantable pulse generator in deep brain stimulation, ACT NEUROCH, 142(5), 2000, pp. 587-589
Citations number
9
Categorie Soggetti
Neurology
Journal title
ACTA NEUROCHIRURGICA
ISSN journal
00016268 → ACNP
Volume
142
Issue
5
Year of publication
2000
Pages
587 - 589
Database
ISI
SICI code
0001-6268(2000)142:5<587:ITSWAM>2.0.ZU;2-R
Abstract
A modified implantable pulse generator (MIPG) for intraoperative lest stimu lation in chronic deep brain stimulation is described. The MIPG can be used for bipolar stimulation with quadripolar electrodes. The device is program med and controlled with a standard console programmer. It can also be used for postoperative test stimulation with externalized electrodes. In our exp erience, the MIPG has several advantages as compared to the screener that i s usually used.