Purpose: To determine the optimal tube current for spiral CT for lung cance
r screening.
Material and Methods: Low-dose spiral CT images from 20 subjects were obtai
ned with 50 mA, 25 mA, 10 mA and 5 mA tube currents. The accuracy in detect
ing 21 nodules measuring 3 to 14 mm in diameter was tested using receiver o
perating characteristics (ROC) to determine the diagnostic accuracy and sta
tistical significance.
Results: Compared with 50 mA, 25 mA showed no statistical difference. A sig
nificant difference was seen between 50 mA and 5 mA images and between 25 m
A and 10 mA images, in the detection of lung parenchymal nodular lesions.
Conclusion: Approximately 25 mA is the optimal tube current for the detecti
on of pulmonary nodules as it allows a reduction of the radiation dose with
out sacrificing diagnostic accuracy.