Rejection of the maternal electrocardiogram in the electrohysterogram signal

Citation
H. Leman et C. Marque, Rejection of the maternal electrocardiogram in the electrohysterogram signal, IEEE BIOMED, 47(8), 2000, pp. 1010-1017
Citations number
29
Categorie Soggetti
Multidisciplinary,"Instrumentation & Measurement
Journal title
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
ISSN journal
00189294 → ACNP
Volume
47
Issue
8
Year of publication
2000
Pages
1010 - 1017
Database
ISI
SICI code
0018-9294(200008)47:8<1010:ROTMEI>2.0.ZU;2-I
Abstract
The electrohysterogram (EHG) signal is mainly corrupted by the mother's ele ctrocardiogram (ECG), which remains present despite analog filtering during acquisition. Wavelets are a powerful denoising tool and have already prove d their efficiency on the EHG, In this paper, we propose a new method that employs the redundant wavelet packet transform. We first study wavelet pack et coefficient histograms and propose an algorithm to automatically detect the histogram mode number. Using a new criterion, we compute a best basis a dapted to the denoising, After EHG wavelet packet coefficient thresholding in the selected basis, the inverse transform is applied. The ECG seems to b e very efficiently removed.