E. Rahiala et T. Tikanoja, NONINVASIVE BLOOD-PRESSURE MEASUREMENTS AND AORTIC BLOOD-FLOW VELOCITY IN NEONATES, Early human development, 49(2), 1997, pp. 107-112
We studied the systolic blood pressure difference between the upper an
d the lower extremities in healthy newborn infants and the effect of t
he isthmic narrowing of the aorta on the possible difference. The bloo
d pressure was measured with an oscillometric blood pressure device fr
om every extremity of 36 healthy infants aged 2-5 days. A Doppler echo
cardiography was performed for each infant to measure the aortic blood
flow velocity in the ascending aorta and in the aortic arch above and
below the isthmic narrowing. The mean blood pressure readings (S.D.)
were the following: the right arm 76.8 (7.3)/48.1 (6.9), the left arm
77.5 (7.4)/51.6 (7.0), the right thigh 77.7 (7.1)/40.7 (5.8), the left
thigh 76.8 (6.4)/39.6 (5.8), the right calf 75.5 (7.1)/46.6 (5.7) and
the left calf 77.1 (8.6)/48.7 (6.7). The aortic blood flow was faster
below the isthmic narrowing of the aorta (1.15+/-0.19 m/s) than in th
e ascending aorta (0.93+/-0.12 m/s) or in the aortic arch above the is
thmus (0.99+/-0.15 m/s). The calculated pressure gradient between the
ascending aorta and aorta below the isthmus was 2.0+/-1.8 mmHg and bet
ween opposite sides of the isthmus 1.5+/-1.2 mmHg. Unlike in childhood
and adolescence, the systolic blood pressure in the lower extremities
of healthy newborn infants is not higher than in the upper extremitie
s. The physiological narrowing of the aortic arch does not explain thi
s phenomenon. If blood pressure measurements are performed on a neonat
e to rule out aortic coarctation, the readings obtained must be interp
reted in respect to normal values in newborns. (C) 1997 Elsevier Scien
ce Ireland Ltd.