An efficient liquid chromatography - Mass spectrometry interface for the generation of electron ionization spectra

Citation
A. Cappiello et al., An efficient liquid chromatography - Mass spectrometry interface for the generation of electron ionization spectra, ANALYT CHEM, 72(16), 2000, pp. 3841-3846
Citations number
12
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICAL CHEMISTRY
ISSN journal
00032700 → ACNP
Volume
72
Issue
16
Year of publication
2000
Pages
3841 - 3846
Database
ISI
SICI code
0003-2700(20000815)72:16<3841:AELC-M>2.0.ZU;2-C
Abstract
The use of a new LC-MS interface (cap-EI), part of a Waters Integrity syste m, capable of generating EI spectra at micro dow rates is presented. The ca p-M interface relies on the production of a fine aerosol by means of a nebu lizer and supported by a nitrogen jet. Sensitivity, response linearity, rep roducibility, and LC compatibility of the interface were thoroughly examine d using testosterone, caffeine, a mixture of antiinflammatory drugs, and 3, 4-dihydroxybenzoic acid as test compounds. The interface is fully compatibl e with LC requirements such as high-water- and/or -buffer-content mobile ph ases. Reproducibility, high sensitivity in scan mode, as well, to produce l ibrary-searchable EI spectra, 2 orders of magnitude linearity, together wit h an intrinsic simplicity of the entire system are the key features of cap- M interface.