J. Chladek et al., High-performance liquid chromatographic assay for the determination of 5-methyltetrahydrofolate in human plasma, J CHROMAT B, 744(2), 2000, pp. 307-313
An isocratic high-performance liquid chromatographic method for the determi
nation of 5-methyltetrahydrofolate (5-MTHF) in human plasma is described. T
he method involves solid-phase extraction of 5-MTHF and p-aminoacetophenon
(an internal standard) using Sep-Pak C-18 cartridges. Separation was achiev
ed with an ODS column using acetonitrile and phosphate buffer supplemented
with octanesulfonic acid (an ion-pairing agent). The pH of the mobile phase
(2.5) was optimal with respect to the mode of detection (fluorescence). Th
e method was validated in the range of 5-MTHF concentrations from 0.0625 mu
mol/l to 4.0 mu mol/l. Within-day and inter-day precision expressed by the
relative standard deviation was less than 8.1% and inaccuracy did not exce
ed 8.7%. The method is specific, accurate and sensitive enough to be used i
n pharmacokinetic studies for the assessment of the systemic availability o
f 5-MTHF after leucovorin administration to patients as a rescue after high
-dose therapy with methotrexate. The limit of detection was 0.17 pmol which
corresponds to a plasma concentration of 1.7 nmol/l. Thus, the assay could
potentially be used for the measurement of 5-MTHF in the range of physiolo
gical concentrations in plasma (5-20 nmol/l). (C) 2000 Elsevier Science B.V
. All rights reserved.