Background: Inflammation associated with biomaterials may contribute to the
failure of glaucoma drainage devices.
Objective: To compare the inflammatory reaction associated with the inserti
on of Krupin silicone, Molteno polypropylene, and Acrosof end plates in the
subconjunctival space of rabbits.
Methods: Similar-sized glaucoma end plates made of 3 different biomaterials
were sutured to the sclera in the superotemporal quadrant of the rabbit ey
e. Thirty eyes of 15 albino New Zealand rabbits were randomly assigned to t
he 3 groups. Conjunctival vascular hyperemia was graded in a masked fashion
among the 3 groups. At the end of 3 weeks, the enucleated eyes were examin
ed histologically and by scanning electron microscopy.
Results: Molteno polypropylene was associated with more inflammation both i
n clinical observations and based on histological grading. Silicone and Acr
osof were associated with less intense inflammation. One polypropylene end
plate was extruded on day 21.
Conclusions: Polypropylene appears to be more inflammatory than silicone. F
lexible biomaterials appear to be less inflammatory than rigid ones.
Clinical Relevance: Bleb failure following glaucoma drainage device implant
ation could be related to the biomaterial-associated inflammation. Choosing
a biomaterial with the least inflammatory potential might enhance the succ
ess rate of the glaucoma drainage device.