Ca. Gonzalez-correa et al., Assessing the conditions for in vivo electrical virtual biopsies in Barrett's oesophagus, MED BIO E C, 38(4), 2000, pp. 373-376
It has previously been shown that it is possible to differentiate between s
quamous and columnar epithelia in rat and resected human tissues using an i
mpedance probe to make in vitro measurements. This probe can be passed down
an endoscope allowing measurements to be made in patients. However, the pr
obe emerges parallel to the oesophageal wall, with little room to manoeuvre
. The conditions of control required to give reliable readings have been in
vestigated The importance of pressure applied and the angle of approach to
the oesophagus was assessed. Pressures in the range 26.6 Pa to 46.3 kPa and
angles in the range 15-90 degrees were considered. In in vitro studies it
was observed that it was possible to obtain consistent readings with pressu
res greater than 2.9 kPa and with angles greater than 15 degrees between th
e probe and the oesophagus. These conditions can be achieved in vivo, and r
eadings obtained from twelve patients are shown (45 readings on normal squa
mous, 34 on Barrett's oesophagus and 22 on stomach). At low frequencies (9.
6-153.2 kHz), a Mann-Whitney test shows a significant difference (p < 0.001
) when comparing the means from squamous and columnar, and also when readin
gs from Barret's and normal gastric epithelia are compared (p < 0.001).