Three-detector setup for positron-lifetime spectroscopy of solids containing Co-60 radionuclide

Citation
J. Cizek et al., Three-detector setup for positron-lifetime spectroscopy of solids containing Co-60 radionuclide, NUCL INST A, 450(2-3), 2000, pp. 325-337
Citations number
11
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
450
Issue
2-3
Year of publication
2000
Pages
325 - 337
Database
ISI
SICI code
0168-9002(20000811)450:2-3<325:TSFPSO>2.0.ZU;2-2
Abstract
A three-detector BaF2 positron-lifetime spectrometer is described. The spec trometer is suited for investigations of the neutron-irradiated reactor pre ssure vessel steels with a high content of Co-60. Both timing and energy in formation about annihilation gamma-rays is used for selection of coincidenc e events. In the triple-coincidence mode, the time resolution of around 220 ps FWHM and the coincidence count rate of 15 s(-1) for a 4 MBq Na-22 posit ron source are simultaneously achieved. Test measurements performed have sh own that in this mode the Co-60 prompt-peak contribution in the positron-li fetime spectrum is suppressed to be below 1% when Co-60 activity of a pair of specimens studied does not exceed the level of 4 MBq. (C) 2000 Elsevier Science B.V. All rights reserved.