Drift velocity monitoring of SDDs using MOS charge injectors

Citation
D. Nouais et al., Drift velocity monitoring of SDDs using MOS charge injectors, NUCL INST A, 450(2-3), 2000, pp. 338-342
Citations number
10
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
450
Issue
2-3
Year of publication
2000
Pages
338 - 342
Database
ISI
SICI code
0168-9002(20000811)450:2-3<338:DVMOSU>2.0.ZU;2-I
Abstract
We report results of drift velocity monitoring in silicon drift detectors, obtained in beam test conditions using MOS charge injectors. We are able to correct velocity variations as small as 0.1% due to temperature variations of the order of 0.1 K and consequently maintain an optimal space resolutio n. (C) 2000 Elsevier Science B.V. All rights reserved.