A quad 500 MHz waveform digitizer with differential trigger for use in themuon g-2 experiment

Citation
S. Dhawan et al., A quad 500 MHz waveform digitizer with differential trigger for use in themuon g-2 experiment, NUCL INST A, 450(2-3), 2000, pp. 391-398
Citations number
2
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
450
Issue
2-3
Year of publication
2000
Pages
391 - 398
Database
ISI
SICI code
0168-9002(20000811)450:2-3<391:AQ5MWD>2.0.ZU;2-L
Abstract
We have developed and tested a 4 channel, 500 MHz, 512 k sample, 8 bit wave form digitizer in a FASTCAMAC module. A novel feature is the implementation of a sparse scan during readout with a programmable differential trigger t o reduce the data to a manageable level. The differential trigger allows re liable detection of pulses with low threshold in the presence of a variable baseline, (C) 2000 Elsevier Science B.V. All rights reserved.