E. Pignoli et al., Dosimetric verification of a commercial 3D treatment planning system for conformal radiotherapy with a dynamic multileaf collimator, PHYS MED BI, 45(8), 2000, pp. N77-N84
The dosimetric accuracy of a 3D treatment planning system (TPS) for conform
al radiotherapy with a computer-assisted dynamic multileaf collimator (DMLC
) was evaluated.
The DMLC and the TPS have been developed for clinical applications where dy
namic fields not greater than 10 x 10 cm(2) and multiple non-coplanar arcs
are required. Dosimetric verifications were performed by simulating conform
al treatments of irregularly shaped targets using several arcs of irradiati
on with 6 MV x-rays and a spherical-shaped, tissue-simulating phantom. The
accuracy of the delivered dose at the isocentre was verified using an ioniz
ation chamber placed in the centre of the phantom. Isodose distributions in
the axial and sagittal planes passing through the centre of the phantom we
re measured using double-layer radiochromic films. Measured dose at the iso
centre as well as isodose distributions were compared to those calculated b
y the TPS. The maximum percentage difference between measured and prescribe
d dose was less than 2.5% for all the simulated treatment plans. The mean (
ISI) displacement between measured and calculated isodoses was, in the axia
l planes, 1.0+/-0.6 mm, 1.2+/-0.7 mm and 1.5+/-1.1 mm for 80%, 50% and 20%
isodose curves, respectively, whereas in the sagittal planes it was 2.0+/-1
.2 mm and 2.2+/-2 mm For 80% and 50% isodose curves, respectively. The resu
lts indicate that the accuracy of the 3D treatment planning system used wit
h the DMLC is reasonably acceptable in clinical applications which require
treatments with several non-coplanar arcs and small dynamic fields.