Quantification of gamma-ray Compton-scatter nondestructive

Citation
Ac. Ho et Ema. Hussein, Quantification of gamma-ray Compton-scatter nondestructive, APPL RAD IS, 53(4-5), 2000, pp. 541-546
Citations number
7
Categorie Soggetti
Multidisciplinary
Journal title
APPLIED RADIATION AND ISOTOPES
ISSN journal
09698043 → ACNP
Volume
53
Issue
4-5
Year of publication
2000
Pages
541 - 546
Database
ISI
SICI code
0969-8043(200010/11)53:4-5<541:QOGCN>2.0.ZU;2-X
Abstract
A method is presented for quantifying information obtained from the indicat ions of a Compton scattering nondestructive testing technique. This is achi eved by formulating a measurement model, which provides a numerical estimat e of the detector response for a given anomaly. Using the measured response function, the location of an anomaly, its size and density are obtained wi th the aid of the model. The process is demonstrated by quantifying indicat ions obtained from experimental measurements for flaws artificially created in an aluminum block. (C) 2000 Elsevier Science Ltd. All rights reserved.