For broad-beam soft X-ray sources, assessment of the quality of image produ
ced by such units is made complex by the low penetration capabilities of th
e radiation. In the present study we have tested the utility of several typ
es of test tool, some of which have been fabricated by us, as part of an ef
fort to evaluate several key image defining parameters. These include the f
ilm characteristic, focal-spot size, image resolution and detail detectabil
ity. The two sources of X-rays used in present studies were the University
of Malaya flash X-ray device (UMFX1) and a more conventional soft X-ray tub
e (Softex, Tokyo), the latter operating at peak accelerating potentials of
20 kV(p). We have established, for thin objects, that both systems produce
images of comparable quality and, in particular, objects can be resolved do
wn to better than 45 mu m. (C) 2000 Published by Elsevier Science Ltd.