A new G-M counter dead time model

Citation
Sh. Lee et Rp. Gardner, A new G-M counter dead time model, APPL RAD IS, 53(4-5), 2000, pp. 731-737
Citations number
14
Categorie Soggetti
Multidisciplinary
Journal title
APPLIED RADIATION AND ISOTOPES
ISSN journal
09698043 → ACNP
Volume
53
Issue
4-5
Year of publication
2000
Pages
731 - 737
Database
ISI
SICI code
0969-8043(200010/11)53:4-5<731:ANGCDT>2.0.ZU;2-M
Abstract
A hybrid G-M counter dead time model was derived by combining the idealized paralyzable and non-paralyzable models. The new model involves two paramet ers, which are the paralyzable and non-paralyzable dead times. The dead tim es used in the model are very closely related to the physical dead time of the G-M tube and its resolving time. To check the validity of the model, th e decaying source method with Mn-56 was used. The corrected counting rates by the new G-M dead time model were compared with the observed counting rat es obtained from the measurement and gave very good agreement within 5% up to 7 x 10(4) counts/s for a G-M tube with a dead time of about 300 mu s. (C ) 2000 Published by Elsevier Science Ltd.