Br. Merz et al., Mechanics of the implant-abutment connection: An 8-degree taper compared to a butt joint connection, INT J O M I, 15(4), 2000, pp. 519-526
Citations number
17
Categorie Soggetti
Dentistry/Oral Surgery & Medicine
Journal title
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS
This paper presents a comparison between the 8-degree Morse Taper and the b
utt joint as connections between an implant and an abutment, Three-dimensio
nal, non-linear finite element models were created to compare the 2 connect
ion principles under equal conditions. The loading configuration was thereb
y modeled according to a test setup actually used for the dynamic long-term
testing of dental implants as required for regulatory purposes, The result
s give insight into the mechanics involved in each type of connection and a
re compared to actual findings with the testing machine, The comparison ind
icates the superior mechanics of conical abutment connections and helps to
explain their significantly better long-term stability in the clinical appl
ication.