Mechanics of the implant-abutment connection: An 8-degree taper compared to a butt joint connection

Citation
Br. Merz et al., Mechanics of the implant-abutment connection: An 8-degree taper compared to a butt joint connection, INT J O M I, 15(4), 2000, pp. 519-526
Citations number
17
Categorie Soggetti
Dentistry/Oral Surgery & Medicine
Journal title
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS
ISSN journal
08822786 → ACNP
Volume
15
Issue
4
Year of publication
2000
Pages
519 - 526
Database
ISI
SICI code
0882-2786(200007/08)15:4<519:MOTICA>2.0.ZU;2-X
Abstract
This paper presents a comparison between the 8-degree Morse Taper and the b utt joint as connections between an implant and an abutment, Three-dimensio nal, non-linear finite element models were created to compare the 2 connect ion principles under equal conditions. The loading configuration was thereb y modeled according to a test setup actually used for the dynamic long-term testing of dental implants as required for regulatory purposes, The result s give insight into the mechanics involved in each type of connection and a re compared to actual findings with the testing machine, The comparison ind icates the superior mechanics of conical abutment connections and helps to explain their significantly better long-term stability in the clinical appl ication.