Zq. Hu et al., A method to separate the transmitting signal and the receiving signal in FM-chirp pulse compression systems, JPN J A P 1, 39(5B), 2000, pp. 3233-3235
We have developed a method of separating the transmitting and the receiving
signals for FM-chirp pulse compression systems for use in intraductal ultr
asonography. The method enables us to remove the unnecessary signals and to
obtain a sufficient delay time. In this paper the following results are de
scribed: 1) A method to separate the transmitting signal and the receiving
signal with a long-time-duration was proposed using a flexible guided line.
2) A guided line suited for intravascular ultrasonography was constructed
using the L(0, 3) mode of the Pocchammer-Chree wave propagating in a tapere
d fused quartz rod. 3) A tentative result was obtained that a transmitting
signal of a long-time-duration could be compressed to give a short pulse. 4
) To use this guided line in intraductal ultrasonography, a method to desig
n an optimum signal is necessary.