Validation of self blood pressure devices

Citation
R. De Gaudemaris et al., Validation of self blood pressure devices, ARCH MAL C, 93(8), 2000, pp. 979-982
Citations number
17
Categorie Soggetti
Cardiovascular & Respiratory Systems
Journal title
ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX
ISSN journal
00039683 → ACNP
Volume
93
Issue
8
Year of publication
2000
Pages
979 - 982
Database
ISI
SICI code
0003-9683(200008)93:8<979:VOSBPD>2.0.ZU;2-Z
Abstract
Validation of automatic devices for self blood pressure measurement is a ne cessity, as the mercury columns are no longer available. The European conse nsus meeting organized by the Group for blood pressure measurement and eval uation (GEM), from French Society of Hypertension held last June 1999 has d efined a new international protocol for clinical validation of these device s. The main parts of this protocol are the following: the blood pressure is me asured at the same arm with two observers simultaneously and then with the automatic device sequentially seven times among 38 patients. The evaluation is interrupted after the 15th patient if the pre-analysis fails defined cr iteria. Technical supports are available :A CD-rom from French society of H TA for the training and certification of observers, a special kit for the e valuation (2 mercury column's and a double stethoscope) and a specific soft ware for data analysis created by the GEM.