Several fluorimetric reactions for the determination of cyanide by mea
ns of flow injection systems were tested, The first reaction is based
on the displacement of 8-hydroxyquinoline-5-sulfonic acid (HQS) from t
he Pd(HQS)(2) complex by cyanide ion and the subsequent reaction of fr
ee HQS with Mg-II to form a fluorescent species, Secondly, the inhibit
ory effect of cyanide on the reaction between iodine and fluorescein w
as also evaluated as a method to quantify cyanide, The third method is
based on the detection of an isoindole derivative formed by the react
ion of o-phthalaldehyde (OPA) or 2,3-naphthalenedialdehyde (NDA) with
glycine in the presence of cyanide, Chemical and hydrodynamic paramete
rs mere optimized for each system and the analytical performance of th
e methods was established, Detection limits of 5 mu g l(-1) (fluoresce
in-iodine method), 0.40 mu g l(-1) (Pd-HQS-Mg method), 0.25 mu g l(-1)
(OPA method) and 0.03 mu g l(-1) (NDA method) were obtained.